CHNS Combustion Analysis

Instrument Information:

The instrument used is a Perkin-Elmer 2400 Series II Analyzer. This instrument can analyze for percent weight of Carbon, Hydrogen, Nitrogen, and Sulfur.

Fees:

CHN Analysis:

(Including one duplicate run)

$30 per sample, for UM shortcode accounts

$68 per sample, for non-UM accounts

Sulfur Analysis:

(Including one duplicate run)

$30 per sample, for UM shortcode accounts

$68 per sample, for non-UM accounts

Sample Information:

Samples can be analyzed in solid, liquid, or oil form. Samples that are oxygen or moisture sensitive can be handled in a dry nitrogen atmosphere in a glove box.

Standard foil capsules, made out of tin, are used to encapsulate solids and viscous oils.

Special crimp-sealed tin pans are used to encapsulate volatile liquids, or are used for solids that might readily lose solvent molecules that are part of the crystal lattice.

Samples are weighed out to an accuracy of 0.001 mg on the Perkin-Elmer AD-4 Auto-Balance.

Samples that are heat sensitive are kept in the freezer until they are analyzed.

If your samples are very unstable, you should contact Carol Carter to schedule a time when they can be analyzed as soon as they are brought to the lab. This can be arranged with prior advanced notification.